U.S. DOE Phantom Library Catalog

Bottle Manikin Absorption (BOMAB) Phantoms

The bottle manikin absorption (BOMAB) phantom consists of 10 high-density polyethylene containers that assemble to represent the ICRP 23 Reference Man human body (see Table 1 for representative container dimensions). It has been used as a standard calibration geometry for over 25 years. The 0.2 to 0.5-cm-thick containers contain a solid polyurethane tissue-substitute with radiation attenuation characteristics matching ICRU 44 average soft tissue. The solid tissue substitute has proven tissue-equivalence and none of the limitations found with water-filled BOMAB phantoms, such as radionuclide plate-out, the potential for leakage, biological growths, and shipment difficulties. The BOMAB polyethylene shells also do not have the large variation in wall thickness that is found with Roto-molded bottle phantoms.

Currently available BOMABs contain one or more of the following isotopes: 40K, 137Cs, 152Eu, 154Eu, 155Eu, 125Sb, 133Ba, and 60Co. Available BOMAB activities are indicated in the Tables BOMAB-nCi and BOMAB-kBq

The BOMAB phantom is used primarily as a calibration standard for Reference Man measurements of homogeneously distributed radioactive materials that emit mid- and high-energy photons. Smaller sizes representing the Reference Female (ICRP 23) and a 4-yr old child are also available.

The tissue substitute used in the BOMAB phantom is a polyurethane-based filler. It has excellent handling characteristics and durability. The plastic matrix contains National Institute of Standards and Technology (NIST) Standard Reference Materials (SRM) or NIST-traceable radioactive materials uniformly distributed throughout the polyurethane. Chemical additives are incorporated to provide the a 40K background in most of the phantoms and to duplicate the transmission characteristics of soft tissues. Specific material constituents and chemical compositions are provided for each phantom.

Table 1. Representative Dimensions of Reference Man BOMAB
Container Quantity Cross section Shape Cross section
Dimensions (cm)
Height
(cm)
Volume
(ml)
1. Head 1 ellipsoid 19x14 20 3525
2. Neck 1 circular 13 dia 10 1030
3. Chest 1 ellipsoid 30x20 40 16970
4. Pelvis 1 ellipsoid 36x20 20 9990
5. Arms 2 circular 10 dia 60 3800 (2)
6. Thighs 2 circular 15 dia 40 6050 (2)
7. Calves 2 circular 12 dia 40 3745 (2)
Total 10     170 58,705
Table 2-nCi. BOMAB Phantom Details
Phantom ID# Radionuclide Calibration Date Activity (nCi)
± Uncertainty
Activity (nCi) on 01Jan2007 Uncertainty Measure Order Info
PL-101 40K - 940±80 940 1sd PNNL
PL-102 137Cs 1/12/90 307±8 207 1sd PNNL
PL-103 137Cs 3/12/92 503.7±8.5 357.7 3sd PNNL
PL-104 154Eu
155Eu
125Sb
1/1/95
1/1/95
1/1/95
994
225
235
±
±
±
58
15
14
386
42.0
11.6
  PNNL
PL-105 133Ba 9/23/92 289.6±2.8 115.2 99%ci PNNL
PL-106 152Eu
40K
6/17/93
-
224.0
115
±
±
1.4
5.75
110.7
115
90%ci
1sd
PNNL
PL-107 137Cs
40K
5/1/93
-
121.3
23
±
±
2.1
1.15
88.4
23
99.7%ci PNNL
PL-108 137Cs
60Co
5/8/96
5/8/96
541.9
537.6
±
±
13.5
5.0
423.4
132.0
  Canada Bureau
PL-109 40K not available 0.16±  0.16   Sandia
Table 2-kBq. BOMAB Phantom Details
Phantom ID# Radionuclide Calibration Date Activity (kBq)
± Uncertainty
Activity (kBq) on 01Jan2007 Uncertainty Measure Order Info
PL-101 40K 4/27/90 34.8±3.0 34.8 1sd PNNL
PL-102 137Cs 1/12/90 11.4±0.30 7.7 1sd PNNL
PL-103 137Cs 3/12/92 18.64±0.31 13.2 3sd PNNL
PL-104 154Eu
155Eu
125Sb
1/1/95
1/1/95
1/1/95
36.78
8.3
8.7
±
±
±
2.1
0.6
0.5
14
1.6
0.4
  PNNL
PL-105 133Ba 9/23/92 10.72±0.10 4.3 99%ci PNNL
PL-106 152Eu
40K
6/17/93
-
8.3
4.255
±
±
0.052
0.21275
4.1
4
90%ci
1sd
PNNL
PL-107 137Cs
40K
5/1/93
-
4.488
0.851
±
±
0.078
0.0426
3.3
0.851
99.7%ci
PNNL
PL-108 137Cs
60Co
5/8/96
5/8/96
20.05
19.89
±
±
0.50
0.19
15.7
4.9
  Canada Bureau
PL-109 40K - 0.0059±  0.0059   Sandia



Phantom ID#: PL-101

Radionuclide: 40K
Reference Man
Tissue Substitute Material: Phantom is filled with ICRU 44 lean muscle tissue-substitute composed of polyurethane, CaCO3, and KOCO(CHOH)2COOH; with a measured variation in activity distribution of <3.4%(n=5).

Phantom ID#: PL-102

Radionuclide: 137Cs
Density of filler: 1.01 g/ml
Reference Man
Tissue Substitute Material:Phantom is filled with ICRU 44 lean muscle tissue-substitute composed of polyurethane and CaCO3, with a measured variation in activity distribution of <3.4%(n=5). No 40K added. The 137Cs SRM was calibrated by counting relative to NIST Reference Point Source #4207B-15.

Phantom ID#: PL-103

Radionuclide: 137Cs
Reference Man
Tissue Substitute Material: Phantom is filled with ICRU 46 adult average soft tissue substitute composed of a elastomer polyurethane, potassium bitartrate, CaCO3, and additives. Filler density is 1.06 g/mL Calculated attenuation coefficients for the tissue substitute are within 2% of ICRU 46 values. Independent elemental analysis of the elastomer results in a composition of: C-64.56%, H-9.52%, N-2.79%, O-23.19%. A nominal amount of 40K is distributed in the tissue material to provide background.

Phantom ID#: PL-104

Radionuclide: 154Eu, 155Eu, 125Sb
Reference Man
Tissue Substitute Material: Phantom is filled with ICRU 44 lean muscle tissue-substitute composed of polyurethane and CaCO3, with a measured variation in activity distribution of 1.1%(n=5). Contains 3.606 g of NIST SRM 4276C-49. No 40K added. The tissue substitute is accurate to within 5% of ICRU 44 lean muscle tissue attenuation coefficients for photons between 100-3000 keV.

Phantom ID#: PL-105

Radionuclide: 133Ba
Reference Man
Tissue Substitute Material: Phantom is filled with ICRU 46 Adult average soft tissue substitute composed of polyurethane, K2CO3, and Bi2O3. Attenuation coefficients for 133Ba photon energies above 75 keV within 0.9% - 6.8% of ICRU 46 values. 40K content approximates ICRP 23 Reference Man.

Phantom ID#: PL-106

Radionuclide: 152Eu
Reference Man
Tissue Substitute Material: Phantom is filled with ICRU 46 Adult average soft tissue substitute composed of polyurethane, potassium bitartrate, and K2CO3 (filler density of 1.022 g/mL). Attenuation coefficients for photon energies are within 2% of ICRU 46 values. 40K content approximates ICRP 23 Reference Man. 88Y, originally present in phantom has decayed out.

Phantom ID#: PL-107

Radionuclide: 137Cs
Canadian BRMD 4-Yr-old Child: Ht. 105 cm., Wt. 18.3 kg
Tissue Substitute Material: Phantom is filled with ICRU 46 average soft tissue substitute composed of polyurethane, potassium bitartrate, and Bi2O3 (filler density 1.03 g/mL). Attenuation coefficients for photon energies are within 1% of ICRU 46 values. 40K content reflects ICRP 23 information. 88Y, originally present in phantom has decayed out.

Phantom ID#: PL-108

Radionuclide: 137Cs, 60Co with 40K background
Reference Female
Total Mass: 55.55 kg. (Active mass 45.46 kg.)
Phantom is filled with PNNL's average soft tissue substitute, composed of an elastomer polyurethane, CaCO3, and K2CO3 for an 81 nCi 40K background contribution (density of filler 1.046 g/mL). The tissue substitute has a total attenuation coefficient within 2% of ICRU 46 average soft tissue. Uniform distribution of 137Cs / 60Co exits in tissue substitute. 40K content reflects ICRP 23 information on Reference Female.

Phantom ID#: PL-109

Blank BOMAB Phantom
Reference Man
Tissue Substitute Material: Phantom is filled with ICRU 46 Average Soft Tissue tissue-substitute composed of polyurethane and 4.31% CaCO3 (density 1.03 g/mL). No 40K added. Trace quantity of 40K from CaCO3 calculated as 0.16 nCi in total phantom. Phantom is useful as a blank measurement for determinations of background spectrum subtraction.